1. Blood cell counters, operating on the principle of conductivity change, which occurs each time a cell passes through an orifice, are generally known as ________________
a) optical method
b) electrical conductivity
c) coulter Counter
d) microscopic method
Answer: c
Clarification: Blood cell counters, operating on the principle of conductivity change, which occurs each time a cell passes through an orifice, are generally known as Coulter Counters. The method was patented by Coulter in 1956 and it forms the basis of several particle counting instruments manufactured by a number of firms throughout the world.
2. Which electrode is placed inside the orifice tube, in coulter counter?
a) Metal electrode
b) Platinum electrode
c) Silicon electrode
d) Magnesium electrode
Answer: b
Clarification: In a coulter counter, a platinum electrode is placed inside the orifice tube and a second electrode is submerged into the beaker containing the cell dilution, creating an electrical circuit between the two electrodes.
3. Which of the following information is not provided by the Coulter Counter?
a) Relative cell size distribution
b) Settings of the threshold level control
c) Relative cell size
d) Mean cell volume
Answer: d
Clarification: The Coulter counters are usually provided with an oscilloscope monitor to display the pulse information, which has passed through the amplifier, and acts as a visible check on the counting process indicating instantaneously any malfunctions such as a blocked orifice. In particular, it provides information regarding (i) relative cell size, (ii) relative cell size distribution, (iii) settings of the threshold level control, and (iv) means to check the performance of the instrument for the reliability of counts.
4. In Coulter counter, for such an aperture, a length of about ______ and flow rate of ______ ml/s would be optimum.
a) 100 u, 0.04
b) 200 u, 0.02
c) 100 u, 0.02
d) 200 u, 0.04
Answer: d
Clarification: Taylor (1970) suggests that an aperture diameter of 100 m would be generally useful. For such an aperture, a length of about 200 μ and flow rate of 0.04 ml/s would be optimum. The aperture can be made using ruby watch jewels bonded to a glass surface.
5. Typically, an aperture of 100 m diameter and 200 m length, separating two solutions of sulphate buffered saline, has a resistance of about 25 kW and capacitance of 120 pF.
a) True
b) False
Answer: b
Clarification: Typically, an aperture of 100 u diameter and 200 m length, separating two solutions of phosphate buffered saline, has a resistance of about 25 kW and capacitance of 120 pF.
6. The required bandwidth of the preamplifier used in cell counter must be?
a) 70 kHz
b) 120 kHz
c) 70 Hz
d) 120 Hz
Answer: a
Clarification: The electronic circuit must have upper frequency response greater than 70 kHz. The preamplifier used in cell counters must be of very low noise preferably having noise voltage less than 2 nA at the required bandwidth of 70 kHz.
7. Which following is not constant for the calibration factor in Coulter counter?
a) Electrolyte resistivity
b) Amplifier gain setting
c) Given aperture size
d) Setting of the threshold level control
Answer: d
Clarification: The calibration factor is constant for given aperture size, electrolyte resistivity and amplifier gain setting. It is used for the conversion of threshold settings to particle volumes or their cube roots to equivalent spherical diameters.
8. What is the diameter of Ragweed pollen?
a) 6 micron
b) 19 micron
c) 14 micron
d) 15 micron