Biomedical Instrumentation Multiple Choice Questions on “Ambulatory Monitoring Instruments”.
1. Where are the electrodes placed for recording foetal electrocardiogram?
a) heart
b) abdomen
c) brain
d) veins
Answer: b
Clarification: Foetal electrocardiogram is recorded by suitably placing the electrodes on the mother’s abdomen and recording the combined maternal and foetal ECG. The maximum amplitude of FECG (R wave) recorded during pregnancy is about 100 to 300 mV. This magnitude is much smaller than in the typical adult ECG which is about 1 mV in the standard lead connection.
2. What is the maximum amplitude of FECG recorded during pregnancy?
a) 100 to 300 mV
b) 300 to 500 mV
c) 500 to 700 mV
d) 700 to 900 mV
Answer: a
Clarification: Foetal electrocardiogram is recorded by suitably placing the electrodes on the mother’s abdomen and recording the combined maternal and foetal ECG. The maximum amplitude of FECG (R wave) recorded during pregnancy is about 100 to 300 mV. This magnitude is much smaller than in the typical adult ECG which is about 1 mV in the standard lead connection.
3. The foetus heart rate is approximately _______ time/s of normal adult foetal heart rate.
a) one
b) two
c) three
d) four
Answer: b
Clarification: The foetal heart rate is computed from the foetal ECG by appropriately shaping the foetal QRS wave. The foetus heart rate is approximately twice that of the normal adult ranging approximately from 110 to 180 bpm. The main problem in processing the foetal heart signals is the poor SNR.
4. Which of the following is periodic noise source in the foetal ECG signal from the maternal abdomen?
a) Amplifier input noise
b) Maternal Muscle noise
c) Fluctuations in electrode polarization potential
d) Maternal ECG
Answer: d
Clarification: The major sources of noise in the foetal ECG signal recorded from the maternal abdomen are (i) amplifier input noise, (ii) maternal muscle noise (EMG), (iii) fluctuations in electrode polarization potential, and (iv) maternal ECG. For practical purposes, the first three of these sources can be considered as random whereas the maternal ECG is a periodic noise source.
5. What is the magnitude of ECG for a typical adult in the standard lead connection?
a) 1 mV
b) 2 mV
c) 3 mV
d) 4 mV
Answer: a
Clarification: Foetal electrocardiogram is recorded by suitably placing the electrodes on the mother’s abdomen and recording the combined maternal and foetal ECG. The maximum amplitude of FECG (R wave) recorded during pregnancy is about 100 to 300 mV. This magnitude is much smaller than in the typical adult ECG which is about 1 mV in the standard lead connection.
6. Which of the following technique is used to measure foetal heart rate?
a) ECG
b) FECG
c) PCG
d) VCG
Answer: b
Clarification: Abdominal FECG processing circuit is used for computing foetal heart rate. After proper placement of the electrodes, the signals are amplified in a preamplifier which provides a very high input impedance (100 MW) and high sensitivity and good common mode rejection ratio (up to 120 dB).
7. What is the CMRR of Foetal electrocardiogram?
a) 0-150 dB
b) 0- 130 dB
c) 0-120 dB
d) 0-200 dB
Answer: c
Clarification: Abdominal FECG processing circuit is used for computing foetal heart rate. After proper placement of the electrodes, the signals are amplified in a preamplifier which provides a very high input impedance (100 MW) and high sensitivity and good common mode rejection ratio (up to 120 dB).
8. What is responsible for most of the common-mode interfering signal?
a) Preamplifier
b) Notch Filter
c) Power Line Hum
d) Band Pass Filter