Basic Statistical Quality Control Questions and Answers on “Gauge and Measurement System Capability Studies – 2”.
1. What of these can be used as a reasonable model for measurement system capability studies? Here y,x and ε denote the observed measurement, true measurement, and the measurement error respectively.
a) y=x-2ε
b) y=x+ε
c) y=x-ε
d) y=x+2ε
Answer: b
Clarification: The reasonable model, which can be used for measurement system capability is expressed as,
y=x+ε
2. The variance of the total observed measurement is expressed by __________
a) (sigma_{Total}^2=sigma_P^2-sigma_{Gauge}^2)
b) (sigma_{Total}^2=sigma_P+sigma_{Gauge}^2)
c) (sigma_{Total}^2=sigma_P^2+sigma_{Gauge})
d) (sigma_{Total}^2=sigma_P^2+sigma_{Gauge}^2)
Answer: d
Clarification: The variance of the total observed measurement uses the value of the sum of the variances of product errors, and the gauge errors. It is calculated by,
(sigma_{Total}^2=sigma_P^2+sigma_{Gauge}^2)
3. The P/T ratio stands for ___________
a) Probability to tolerance ratio
b) Precision to time ratio
c) Probability to total ratio
d) Precision to tolerance ratio
Answer: d
Clarification: The P/T ratio is used in the measurement system capability analysis. In P/T ratio, P/T refers to Precision to tolerance ratio.
4. What is the value of the P/T ratio?
a) (frac{P}{T}=frac{1.5khat{σ_p}}{USL-LSL})
b) (frac{P}{T}=frac{khat{σ}_{gauge}}{USL-LSL})
c) (frac{P}{T}=frac{2khat{σ_p}}{USL+LSL})
d) (frac{P}{T}=frac{khat{σ}_{gauge}}{USL+LSL})
Answer: b
Clarification: The P/T ratio is calculated for the evaluation of the gauge capability. It uses the value of the (hat{σ_{gauge}}) The value of P/T ratio is given by,
(frac{P}{T}=frac{khat{σ}_{gauge}}{USL-LSL})
5. If the number of standard deviations between the usual natural tolerance limits of a normal distribution, what is the value used for k in the P/T ratio?
a) 5.15
b) 8
c) 6
d) 5.60
Answer: c
Clarification: The value k=6 in the P/T ratio corresponds to the number of standard deviations, between the usual natural tolerance limits for a normal distribution.
6. For a process, which has, USL and LSL equal to 60, and 5 respectively, and the value of (hat{σ}_{gauge}) = 0.887, what will be the value of P/T ratio when k=6?
a) 0.087
b) 0.077
c) 0.067
d) 0.097
Answer: d
Clarification: We know that,
(frac{P}{T}=frac{khat{σ}_{gauge}}{USL-LSL})
Putting the values in the question, we get P/T=0.097.
7. Which of these indicate an adequate measurement system?
a) (frac{P}{T}≤0.1)
b) (frac{P}{T}≤0.5)
c) (frac{P}{T}≥0.1)
d) (frac{P}{T}=0.3)
Answer: a
Clarification: If the value of P/T ratio is lesser than or equal to 0.1, the measurement system used is predicted to be adequate for the selected process.
8. The options are the P/T ratios for different measurement systems. Which of these shows an adequate measurement system?
a) 0.21
b) 0.13
c) 0.18
d) 0.06
Answer: d
Clarification: A P/T ratio lesser than 0.1 indicates an appropriate measurement system for any process. So here, 0.06<0.1, so it is an example of acceptable measurement systems.
9. The cause of calling a measurement system adequate because it has P/T ratio lesser than 0.1, is ____________
a) A measurement device should be calibrated in units one-tenth large as the accuracy required in final measurement
b) A measurement device should be calibrated in units one-third large as the accuracy required in final measurement
c) A measurement device should be calibrated in units one-fourth large as the accuracy required in final measurement
d) A measurement device should be calibrated in units three-tenth large as the accuracy required in final measurement